Temperature Sensor
Location / Function
During the Fill Mode...
e temperature sensor is not monitored.
During the Heat-Up / Sterilization Modes...
e temperature sensor continually monitors the chamber
temperature and transmits this information to the PC board.
e PC board turns the heating element ON / OFF based on
the readings from the temperature sensor.
During the Vent Mode...
e temperature sensor continually monitors the chamber
temperature and transmits this information to the PC board.
During the Dry Mode...
e temperature sensor continually monitors the chamber
temperature and transmits this information to the PC board.
If the temperature exceeds 240°F (115°C), the PC board stops the current
ow to the heating element until the temperature drops.
* Service Tip
Residue build-up can cause inaccurate temperature / pressure readings.
Clean the sensor probe with Speed-Clean and distilled water.
Sensor Probe*