Door Switch
Location / Function
NOTE:
When the door is OPEN, the door switch is untripped / OPEN.
When the door is CLOSED, the door switch is tripped / CLOSED.
During the Fill / Heat-Up / Sterilization Modes...
When a cycle is initiated, the PC board monitors the status of the door
switch.
If an open door is detected, the cycle will not start. If the door switch opens
during a cycle, the cycle will be terminated and the corresponding error code
will appear in the display.
During the Vent / Drying Modes...
e door switch is not monitored.
WARNING
The door switch should remain untripped
until door latch is completely closed.