High-Limit Thermostats - continued
Replacement
Step 1: Remove thermostats.
A) Drain water from reservoir.
B) Disconnect wires from heating element and thermostats.
D) Remove nuts, lockwashers, and brass washers.
E) Remove bracket and thermostats.
Equipment Alert
Do not overtighten nuts!
Torque must not exceed 25 ft/lbs (34 N•m).
Step 2: Install thermostats.
A) Inspect heating element gaskets for damage.
Replace gaskets if necessary.
B) Place thermostats and bracket in position.
C) Install brass washers, lockwashers, and nuts.
D) Connect wires to thermostats and heating element.
Brass Washer
Lockwasher
Nut
Wire Connections
White Jumper wire .................. between ermostats
Yellow wire ............................... ermostat
Red wire .................................... ermostat
Brown wire ............................... Heating Element
Brown / White wire ................. Heating Element
Equipment Alert
Thermostats must be tight against
chamber wall when assembled.