English-113
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Main PC Board
Location / Function
During all modes...
e Main PC Board controls all of the electronic components of the steril-
izer. During operation, the pressure sensor monitors the chamber condi-
tions to maintain the parameters for the selected cycle.
e two fuses (F1 & F2) protect the circuitry from excessive current draw.
If either fuse is faulty, the unit will not operate.
e SW1 switches are used for Service Diagnostics, and to adjust the
display to metric units.
SW1 switches
Fuses
(F1 / F2)
midmark.com