EasyManua.ls Logo

Midmark M11 - Using a Pressure Gauge

Default Icon
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English-73
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Using a Pressure Gauge
MA617801i
Note
To test chamber pressure, a Pressure Gauge Harness is available (002-0372-00).
Step 1: Connect the pressure gauge.
A) Cut cable tie, then disconnect tubing from PC board.
B) Connect pressure gauge harness as shown.
Step 2: Verify correct pressure display.
A) Run a sterilization cycle.
B) The pressure shown on the gauge should match the
pressure on the display panel.
Step 3: Remove gauge / Connect tubing.
A) Disconnect pressure gauge harness.
B) Connect tubing to PC board.
C) Secure tubing with high-temp cable tie.
midmark.com
Style A

Table of Contents

Other manuals for Midmark M11

Related product manuals