High-Limit Thermostats
Location / Function / Testing
During all Modes...
Line voltage continually ows thru the normally closed contacts of the two
high-limit thermostats. is circuit powers all of the line voltage components,
except for the fan system.
If the temperature at either of the thermostats exceeds 450°F (+/- 25°) / 232°C
(+/- 14°), the thermostat contacts open. is interrupts power, and terminates
the cycle. [An error code will appear on the display].
e thermostat contacts reset to the closed position at approximately 325°F /
163°C.
Note: Thermostats must be tested at room temperature.
High-Limit Thermostat: Continuity Test
A) Disconnect sterilizer power cord.
B) Disconnect wires from thermostat.
C) Set meter to ‘200 Ω’
D) Place meter probes on thermostat terminals.
Meter Reading Required Action
approx. 0.0
Thermostat is good
OL Replace thermostat