Service Diagnostics
e Service Diagnostics feature allows you to view recent error codes and test the sterilizer’s major components without running a complete cycle. e Service Diag-
nostics tests should always be done before replacing any major component.
Activating Service Diagnostics
To activate Service Diagnostics...
A) Disconnect sterilizer power cord.
B) Remove RH side panel.
C) Move switch #1 on SW1 block to ON.
D) Reconnect power cord.
E) Press Start button.
Note: Go to “Test Selection Screen” page for next step.
Caution
This operation requires power to be connected to the unit with
panels removed. Use caution when performing this procedure.
To return to normal operating mode...
A) Disconnect power cord.
B) Move switch #1 (SW1 block) to OFF.
C) Reconnect power cord.