Main PC Board - continued
SW1 Switch Settings
e eight SW1 switches are set to the OFF position when shipped from the factory.
ese switches are used when:
•ActivatingtheServiceDiagnosticsMode
•ConguringthePCBoard(requiredwhenboardisreplaced)
•Changingthedisplaytometricunits(Celsius/kPa)
ISA102401i
Switch 1
Switch 2
Switch 3
Switch 4
Switch 5
Switch 6
Switch 7
Switch 8
1
2
3
4
5
6
7
8
Switch Positions
OPTIONS
SW-1
ON
Turn ON for Service
Diagnostics Mode
Turn ON to configure PC Board
Not Used, leave OFF
Not Used, leave OFF
Not Used, leave OFF
Not Used, leave OFF
OFF - English Units
ON - Metric Units
Not Used, leave OFF
To activate the Service Diagnostics Mode...
Move Switch 1 to ON position.
Note: This switch must be turned OFF to resume normal operation.
To change the display to metric units...
Move Switch 7 to ON position.
To configure the PC Board...
Move Switch 2 to ON position.
Note: This procedure changes the software settings for
chamber size (ex. M9 / M11), and the door opening
function (ex. M9 / M9D).