English-77
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Service Diagnostics - continued
I/O Test - continued
Door Open Test
Note: This test should be done with the door closed.
Press the Start button.
This energizes the Door Motor System.
[The door should open after approx. 15 seconds.
This indicates the PC board and door motor are functioning properly.]
Press the Stop button for the next test.
Steam Heater Test
Equipment Alert
Do not run this test more than twice without allowing unit to cool.
Doing so may cause the sterilizer to overheat.
Press the Start button.
This energizes the Heating Element.
[The heating element should heat up for approx. 15 seconds, then shut off.
This indicates the PC board and heating element are functioning properly.]
Press the Stop button for the next test.
Door Motor System
Heating Element