Checking for Pressure Leaks
Component Check Correction
Door / Dam Gaskets Check for water leaking around door. Inspect / clean gaskets.
Replace gasket(s) if necessary.
Vent Valve Check for water leaking from condensing coil spout. Clean / replace vent valve.
Fill Valve Check for water leaking back into reservoir thru the ll line.
Look for bubbles coming from bottom of reservoir.
Clean / replace ll valve.
Air Valve Check for excessive steam coming thru condensing coil spout.
NOTE: During the HEAT & VENT modes, it is normal for steam to be exhausted thru the spout.
Clean / replace air valve.
Pressure Relief Valve Check for water / steam leakage from beneath the rear of the sterilizer. Refer to: Pressure Relief Valve Test.
Replace valve if necessary.
Pressure Sensor Hose Check for steam leakage at pressure sensor hose connections. Secure pressure sensor hose connection
with high temperature cable ties.
Door Gasket
Dam Gasket
Fill Valve
Vent Valve
Pressure Sensor Hose
Connections
Air Valve
Pressure Relief Valve
Condensing Coil Spout
Check for bubbles here.
Check for water / steam leakage here.
WARNING
Do not attempt to adjust, modify,
or alter in any manner, any part of
the pressure vessel. Serious injury and/or
damage to the unit may result.