EasyManua.ls Logo

Midmark M11 - Fan System; Fan Thermostat

Default Icon
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English-98
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: 02/22/13]
Fan System
Location / Function
NOTE:
e fan may run continuously when running consecutive cycles.
During all Modes...
When power is supplied to the main PC board, line voltage continuously ows to the
fan thermostat.
e fan thermostat controls the ON/OFF function of the fan. When the temperature
(at the thermostat) is below 130°F (54°C), the fan thermostat contacts are open (no
current to the fan is OFF). When the temperature reaches 130°F (54°C), the fan
thermostat contacts close (current ows to the fan is ON).
When the temperature drops to approx. 100°F (38°C), the contacts of the fan ther-
mostat open and the fan stops running.
Fan Thermostat
Fan
midmark.com