Error Code Explanation
If an electronic malfunction is detected during a cycle, a numeric error code will appear on the display panel.
Each digit in the error code provides information about the problem that occurred.
Example:
First Digit = Where
e rst digit indicates the component
or system where the problem occurred.
(example: 3 = Door Switch)
Second Digit = What
e second digit indicates what problem
or symptom was detected.
(example: 8 = Open)
Third Digit = When
e third digit indicates when the problem was detected.
(example: 2 = Fill Mode)
The table below cross-references the numeric error code with the Component, Problem, and Mode.
First Digit
(Component)
Second Digit
(Problem)
Third Digit
(Mode)
0 = General System 0 (not used) 0 = Power-Up Mode
1 = Stop Button 1 = Power Loss 1 = Select Cycle
2 = Water Level Sensor 2 = Closed 2 = Fill Mode
3 = Door Switch 3 = Low 3 = Heat-Up Mode
4 (not used) 4 = High 4 = Sterilizing Mode
5 = Temperature Sensor 5 (not used) 5 = Vent
6 = Pressure Sensor 6 = Hardware 6 = Door To Open
7 (not used) 7 = Over Limit 7 = Dry
8 (not used) 8 = Open 8 (not used)
9 = High Limit ermostat 9 (not used) 9 (not used)
Trouble Shooting Chart
C010: (System Power Loss)
C060: (System Hardware)
C100 Series: (Stop Key)
C230 Series: (Water Low)
C320 Series: (Door Closed)
C380 Series: (Door Open)
C530 Series: (Steam Temp. Low)
C540 Series: (Steam Temp. High)
C560 Series: (Steam Temp. Hardware)
C570 Series: (Steam Temp. Over limit)
C630 Series: (Pressure Low)
C640 Series: (Pressure High)
C660 Series: (Pressure Hardware)
C670 Series: (Pressure Over limit)
C980 Series: (Hi-Limit Open)