Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Error Code:
C010
C010: POWER MODE
SYSTEM PWR LOSS
ITEMS NOT STERILE
PUSH STOP TO RESTART
Power interruption.
(Unit lost power for a few seconds during
Fill, Heat-Up, or Sterilization Mode.)
Check power cord connections.
(Power interruption could be due to
an electrical storm, brown out, etc.)
Secure power cord connections.
(At wall outlet & at sterilizer receptacle)
Unplug unit for 60 seconds, then plug
unit into a dedicated, properly rated circuit.
Loose / Damaged wire connections. Check AC terminals and connections
at J5 on main PC board
Secure / Repair loose or damaged
connections.
Power cord/receptacle damaged.
Inspect condition of power cord/ receptacle. Replace power cord/ receptacle.
Ribbon cable loose or damaged. If unit is beeping/clicking with no display.
(or display cuts out)
Secure/replace ribbon harness.
Diplay PC compromised. Check for damaged pins/connection to ribbon
cable.
Replace diplay PC board.
Error Code will not reset. Unit must be unplugged for 60 seconds. Replace Main PC board.
Error Code:
C060
C060: POWER MODE
SYSTEM HARDWARE
ITEMS NOT STERILE
UNPLUG/RE-PLUG UNIT
Power interruption.
(Unit lost power for a few seconds during
Fill, Heat-Up, or Sterilization Mode.)
Check power cord connections.
(Power interruption could be due to
an electrical storm, brown out, etc.)
Unplug unit for 60 seconds, then plug
unit into a dedicated, properly rated circuit.
Check supply voltage. Contact an electrician to modify voltage.
Loose / Damaged wire connection. Check AC terminals and connections
at J5 on main PC board.
Secure / Repair loose or damaged
connections.
Power cord/receptacle damaged.
Inspect condition of power cord/ receptacle. Replace power cord/ receptacle.
Ribbon cable loose or damaged. If unit is beeping/clicking with no display.
(or display cuts out)
Secure/replace ribbon harness.
Diplay PC compromised. Check for damaged pins/connection to ribbon
cable.
Replace diplay PC board.
Error Code will not reset. Unit must be unplugged for exactly 60 seconds. Replace Main PC board.
Error Code:
C099
C099: MISC MODE ??? ???
PUSH STOP TO RESTART
Code was generated during factory testing No check necessary. Press STOP button and initiate a new cycle
Error Code:
C102
C102: FILL MODE
STOP PRESSED
STOP button was pressed. No Check Necessary. Press STOP button and initiate a new cycle.
Touch pad malfunctioning.
Perform Key test in service diagnostic.
Refer to: Test Selection Screen
Replace touch pad.
Error Code:
C103
C103: HEATUP MODE
STOP PRESSED
STOP button was pressed. Check if chamber pressure and temperature
has dissipated, if not it may be necessary to
allow chamber to cool.
Press STOP button and initiate a new cycle.
Touch pad malfunctioning.
Perform Key test in service diagnostic.
Refer to: Test Selection Screen
Replace touch pad.
Error Code:
C104
C104: STERILIZE MODE
STOP PRESSED
STOP button was pressed. Check if chamber pressure and temperature
has dissipated, if not it may be necessary to
allow chamber to cool.
Press STOP button and initiate a new cycle.
Touch pad malfunctioning.
Perform Key test in service diagnostic.
Refer to: Test Selection Screen
Replace touch pad.
Error Code:
C105
C105: VENT MODE
STOP PRESSED
STOP button was pressed. Check if chamber pressure and temperature
has dissipated, if not it may be necessary to
allow chamber to cool.
Press STOP button and initiate a new cycle.
Touch pad malfunctioning.
Perform Key test in service diagnostic.
Refer to: Test Selection Screen
Replace touch pad.