Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Water Leaking Under Unit.
Water Puddling under door. Sterilizer is not level.
(remove tray / trays)
Verify support surface is level.
Place sterilizer on a level support surface.
Verify water on all sides of chamber
are equal.
Place sterilizer on a level support surface.
Gap between gasket housing and
dam gasket.
Check if there is a gap in between the
gasket housing and dam gasket.
Replace dam gasket.
Weak or worn gasket kit. Check for loose / worn door and dam
gaskets or water leaking around door.
Verify door and dam gaskets are installed
properly. (Replace gasket kit 1x year)
Clean / Replace gaskets.
Gasket housing defective. Check distence between gasket housing pins. Replace gasket housing.
Heating element gasket leaking. Check for corrosion from moisture leaking
at heating element terminals or thermostat
terminals.
Replace heating element gaskets.
Loose/missing reservoir clamp. Check if hose clamp at the reservoir is
connection to the drain tube is loose,
disconnected or damaged.
Tighten/replace clamp.
Reservoir /tubing leaking. Check reservoir and tubing from the reservoir to
the manifold assembly for leaks.
Repair/replace reservoir or tubing.
Manifold block leaking. Check manifold block assembly for leaks Replace Fill/Vent Manifold assembly