English-75
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Service Diagnostics - continued
Test Selection Screen
Press the Start button to initiate the I/O Test.
is test allows you to energize the air valve, vent valve, ll
valve, door motor, and heating element independently without
running a cycle.
is test also displays the temperature / pressure inside the
chamber, and the status of the high-limit thermostats, door
switch, and the water level sensor.
Press the Stop button to show the last ve error codes dis-
played on the unit.
Press the Handpieces button to initiate the Keytest.
is test allows you to check the functionality of the
buttons on the touch pad.