[Revised: mo/dd/yr]
Service Diagnostics - continued
I/O Test - continued
Sensor Status
The display shows the status of the High-Limit Thermostats, the
Door Switch, and the Water Level Sensor.
[If the display reading indicates a malfunction, test the corresponding component]
Press the Stop button for the next test.
High-Limit Thermostats
Status should always be CLOSED.
OPEN, indicates malfunctioning thermostat(s), or
that the unit has overheated
Door Switch
Status should correctly reect
the position of the door.
Water Level Sensor
Status should reect the water level in the chamber.
If water is contacting the sensor, status should be
FULL. If not, EMPTY.
Chamber Temperature
[Verify w/ thermometer]
Chamber Pressure
With the door open, display
should show: 0.0 PSI (0.0 kPa)
Chamber Temperature / Pressure
The display shows the temperature / pressure inside the chamber.
Press the Stop button to return to the Test Selection Screen.