Service Diagnostics - continued
I/O Test - continued
Sensor Status
The display shows the status of the High-Limit Thermostats, the
Door Switch, and the Water Level Sensor.
[If the display reading indicates a malfunction, test the corresponding component]
Press the Stop button for the next test.
High-Limit Thermostats
Status should always be CLOSED.
OPEN, indicates malfunctioning thermostat(s), or
that the unit has overheated
Door Switch
Status should correctly reect
the position of the door.
Water Level Sensor
Status should reect the water level in the chamber.
If water is contacting the sensor, status should be
FULL. If not, EMPTY.
Chamber Temperature
[Verify w/ thermometer]
Chamber Pressure
With the door open, display
should show: 0.0 PSI (0.0 kPa)
Chamber Temperature / Pressure
The display shows the temperature / pressure inside the chamber.
Press the Stop button to return to the Test Selection Screen.