English-80
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Service Diagnostics - continued
Keytest
Keytest
Press the Start button.
[When the designated button is pressed, you will hear a single
“beep”, then the test will advance to the next button. This indicates
the button is functioning properly].
Keytest
Press the Stop button.
Keytest
Press the Handpieces button.
[Continue for all remaining buttons].