English-89
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
SA172802i
Pressure Relief Valve - continued
Replacement
To replace the pressure relief valve...
A) Pull pressure relief handle to purge pressure from chamber.
B) Disconnect relief valve tubing, then separate valve enclosure.
C) Remove pressure relief valve.
D) Install new pressure relief valve*.
E) Position valve washer between top of enclosure and valve ring,
then install LH / RH valve enclosure.
F) Connect relief valve tubing. Trim tubing if necessary to prevent
contact with support surface. (M9: 13 3/4” / M11: 16 1/4”)
*Note: Pressure relief valves have thread sealant on them to prevent leakage.
Valve Enclosure (LH)
Valve Enclosure (RH)
Washer