Fill / Vent Valve
Location / Function
Fill Valve
During the Fill Mode...
Voltage is supplied to the ll valve. is causes the valve to open, allowing
water to ow into the chamber.
When the water in the chamber contacts the water level sensor, the PC
board stops the current ow to the ll valve. is allows the valve to close,
stopping the ow of water into the chamber.
During the Heat-Up / Sterilization / Vent / Dry Modes...
ere is no current ow to the ll valve. e valve is closed.
Vent Valve
During the Fill / Heat-Up / Sterilization Modes...
Voltage is supplied to the vent valve. is causes the valve to close so that
pressure can build in the chamber.
During the Vent Mode...
e PC board stops the current ow to the vent valve. is allows the valve
to open, releasing steam / pressure from the chamber.
During the Dry Mode...
ere is no current ow to the vent valve. e valve is open.
Vent Valve