EasyManua.ls Logo

Midmark M11 - I;O Test; Air Valve Test; Vent Valve Test; Fill Valve Test

Default Icon
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English-76
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Service Diagnostics - continued
I/O Test
Air Valve Test
Press the Start button.
This energizes the Air Valve, causing it to open.
Press the Start button again to close the valve.
[You should hear a “click” when the valve opens / closes.
This indicates the PC board and valve are functioning properly.]
Press the Stop button for the next test.
Vent Valve Test
Press the Start button.
This energizes the Vent Valve, causing it to close.
Press the Start button again to open the valve.
[You should hear a “click” when the valve opens / closes.
This indicates the PC board and valve are functioning properly.]
Press the Stop button for the next test.
Fill Valve Test
Equipment Alert
The door switch must be tripped when testing the Fill Valve. Close the door
or manually trip the switch. The water level sensor does not function during this
test. The chamber will overflow if the valve is left open too long.
Press the Start button.
This energizes the Fill Valve, causing it to open.
Press the Start button again to closes the valve.
[Water will flow into the chamber when the valve opens.
This indicates the PC board and valve are functioning properly.]
Press the Stop button for the next test.
Air Valve
Fill Valve
Vent Valve
midmark.com
Style A

Table of Contents

Other manuals for Midmark M11

Related product manuals