Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Error Code:
C570
Note: When a code ends in “0”, it means
the unit had a power interruption during
operating conditions. The operator may
have unplugged the unit when a previous
error code was displayed.
C570: POWER UP MODE
STEAM TEMP OVERLIM
UNPLUG / REPLUG UNIT
Steam temperature exceeds 284°F
(140°C) during POWER UP mode.
Power interruption.
(Unit lost power for a few seconds during
Fill, Heat-Up, or Sterilization Mode.)
Check power cord connections.
(Power interruption could be due to an
electrical storm, brown out, etc.)
Secure power cord connections.
(At wall outlet & sterilizer receptacle)
The operator may have unplugged
the unit when a previous error code was
displayed.
Check previous error codes. Refer to: Service
Diagnostics (Recall Errors)
Refer to the error code(s) in this
troubleshooting chart.
Temperature sensor malfunctioning. Check for residue build up on temperature
sensor probe.
Clean sensor probe with Speed-Clean,
distilled water, and abrasive pad.
Check that nothing is touching or crowding
the temperature sensor.
(Example: trays, pouches, cassettes, etc.)
Reposition items so nothing touches or
crowds the temperature sensor.
Perform Temperature Sensor
Resistance Test. Allow sensor to cool to
room temperature before testing.
Acceptable range 1.07-1.1K Ω.
Refer to:
Temperature Sensor Resistance Test
Replace temperature sensor.
Perform Temperature Sensor Supply
Voltage Test. Disconnect sensor
harness from J12 on PC board.
On pins 1 & 2 acceptable range is
4.0 - 6.0 VDC. Refer to:
Temperature Sensor Supply Voltage Test
Replace Main PC board.
Sterilizer is unlevel.
(remove tray / trays)
Verify support surface is level. Place sterilizer on a level support surface.
Verify water on all sides of chamber
are equal.