English-33
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: 8/15/17]
Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Error Code:
C571
.
C571: SELECT MODE
STEAM TEMP OVERLIM
UNPLUG / REPLUG UNIT
Steam temperature exceeds 284°F
(140°C) during SELECT mode.
Temperature sensor malfunctioning. Check for residue build up on temperature
sensor probe.
Clean sensor probe with Speed-Clean,
distilled water, and abrasive pad.
Check that nothing is touching or crowding
the temperature sensor.
(Example: trays, pouches, cassettes, etc.)
Reposition items so nothing touches or
crowds the temperature sensor.
Perform Temperature Sensor
Resistance Test. Allow sensor to cool to
room temperature before testing.
Acceptable range 1.07-1.1K Ω.
Refer to:
Temperature Sensor Resistance Test
Replace temperature sensor.
Perform Temperature Sensor Supply
Voltage Test. Disconnect sensor
harness from J12 on PC board.
On pins 1 & 2 acceptable range is
4.0 - 6.0 VDC. Refer to:
Temperature Sensor Supply Voltage Test
Replace Main PC board.
Sterilizer is unlevel.
(remove tray / trays)
Verify support surface is level. Place sterilizer on a level support surface.
Verify water on all sides of chamber
are equal.
Error Code:
C572
C572: FILL MODE
STEAM TEMP OVERLIM
UNPLUG / REPLUG UNIT
Steam temperature exceeds 284°F
(140°C) during FILL mode.
Temperature sensor malfunctioning. Check for residue build up on temperature
sensor probe.
Clean sensor probe with Speed-Clean,
distilled water, and abrasive pad.
Check that nothing is touching or crowding
the temperature sensor.
(Example: trays, pouches, cassettes, etc.)
Reposition items so nothing touches or
crowds the temperature sensor.
Perform Temperature Sensor
Resistance Test. Allow sensor to cool to
room temperature before testing.
Acceptable range 1.07-1.1K Ω.
Refer to:
Temperature Sensor Resistance Test
Replace temperature sensor.
Perform Temperature Sensor Supply
Voltage Test. Disconnect sensor
harness from J12 on PC board.
On pins 1 & 2 acceptable range is
4.0 - 6.0 VDC. Refer to:
Temperature Sensor Supply Voltage Test
Replace Main PC board.
Sterilizer is unlevel.
(remove tray / trays)
Verify support surface is level. Place sterilizer on a level support surface.
Verify water on all sides of chamber
are equal.