Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Error Code:
C670
Note: When a code ends in
“0”, it means the unit had a
power interruption during
operating conditions. The
operator may have unplugged
the unit when a previous error
code was displayed.
C670: POWER UP MODE
PRESSURE OVERLIM
UNPLUG / REPLUG UNIT
Gauge pressure inside chamber is
greater than 34.8 psig (240 kPa)
during POWER UP mode.
Power interruption
(Unit lost power for a few seconds during
Fill, Heat-Up, or Sterilization Mode.)
Check power cord connections.
(Power interruption could be due to
an electrical storm, brown out, etc.)
Unplug unit for 60 seconds, then plug
unit into a dedicated, properly rated circuit.
Check supply voltage. Contact an electrician to modify voltage.
The operator may have unplugged
the unit when a previous error code was
displayed.
Check previous error codes. Refer to: Service
Diagnostics (Recall Errors)
Refer to the error code(s) in this
troubleshooting chart.
Vent valve. Inspect vent valve.
(Clean valve, stretch spring)
Clean / Replace vent valve.
With the sterilizer in the Vent Mode...
Perform: Vent Valve Testing
(The valve is normally open, and
should not be energized during
the Vent mode).
Pressure tubing at incorrect angle. Check tubing for restrictions and
ensure tubing is angled down and
away from the PC Board.
Adjust pressure tubing to correct angle.
PC Board malfunctioning.
Perform: Heating Element Supply
Voltage Test (With harness attached
to leads should have line voltage in
test mode)
Replace Main PC board.
Perform: Main PC Board Relay Test
(Check TP1 & TP2 for an acceptable range: 10
to 14 VDC)
Perform: Main PC Board Pressure
Transducer Voltage Test
(Acceptable range: 4.0 to 6.0 VDC)
Error Code:
C671
t.
C671: SELECT MODE
PRESSURE OVERLIM
UNPLUG / REPLUG UNIT
Gauge pressure inside chamber is
greater than 34.8 psig (240 kPa)
during SELECT mode.
PC Board malfunctioning.
Perform: Heating Element Supply
Voltage Test (With harness attached
to leads should have line voltage in
test mode)
Replace Main PC board.
Perform: Main PC Board Relay Test
(Check TP1 & TP2 for an acceptable range: 10
to 14 VDC)
Perform: Main PC Board Pressure
Transducer Voltage Test
(Acceptable range: 4.0 to 6.0 VDC)