English-99
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: mo/dd/yr]
Fan System - continued
Testing
To test the fan...
A) Disconnect two wires from fan.
B) Set meter to ‘200 Ω’
C) Place meter probes on fan terminals.
Meter Reading Required Action
Continuity (
Ω)
Fan is OK
‘0’ or no reading Replace fan
To test the fan thermostat...
A) Disconnect two wires from fan thermostat.
B) Set meter to ‘200 Ω’
C) Place meter probes on thermostat terminals.
Meter Reading Required Action
OL Fan thermostat is OK
anything other than OL Replace fan thermostat
midmark.com