Water Level Sensor - continued
Replacement
MA665200i
Service Tip:
Residue build-up on the sensor disk can cause malfunctions during the Fill mode.
Clean the sensor disk with Speed-Clean and an abrasive pad, then dry disk thoroughly.
Step 1: Remove water level sensor.
A) Disconnect wire from sensor terminal.
B) Remove nut, terminal, and compression nut.
C) (From inside chamber) Remove sensor, spacer, and teflon tube.
Equipment Alert
Do not overtighten compression nut!
Tighten 1-1/4 turns past finger-tight.
Step 2: Install water level sensor.
A) Install teflon tube and spacer onto sensor.
B) Insert sensor thru hole in chamber.
C) Install compression nut.
D) Install terminal and nut.
E) Perform water level sensor test to be sure
the sensor is not grounded.
F) Connect wire to sensor terminal.
Nut
Terminal
Compression Nut
Teon Tube
Spacer
Sensor
midmark.com