English-116
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: 9/26/17]
Main PC Board - continued
Replacement
MA8745i
1
2
3
4
5
6
7
8
OPTIONS
SW-1
ON
Step 1: Remove main PC board.
A) Tag / disconnect all wire harnesses from PC board.
B) Cut cable tie, then disconnect pressure sensor tubing.
C) Remove two nuts and PC board / bracket.
Step 2: Install main PC board.
A) Install PC board / bracket, secure with two nuts.
B) Connect pressure sensor tubing, then secure with
high-temperature cable tie.
C) Connect all wire harnesses to PC board.
D) Unplug the power cord.
E) Move switch 2 on the SW1 block to ON.
F) Reconnect power cord.
Step 3: Configure the main PC board.
A) Adjust the PC board configuration by following
the prompts on the display panel.
Use the < + > < - > buttons to adjust settings.
Press the < P > button when finished.
CHAMBER DIAMETER:
M9/M9D ...................... 9 INCH
M11/M11D ..................11 INCH
FEATURE SET:
M9/M11 ....................... FULLY FEATURED
M9D/M11D .................DEFEATURED
Pressure Sensor Tubing
Step 4: Return to normal operating mode.
A) Disconnect power cord.
B) Move switch #2 (SW1 block) to OFF.
C) Reconnect power cord.