EasyManua.ls Logo

Midmark M11 - Page 116

Default Icon
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English-116
004-1032-00
TP210 Rev. A
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
[Revised: 9/26/17]
Main PC Board - continued
Replacement
MA8745i
1
2
3
4
5
6
7
8
OPTIONS
SW-1
ON
Step 1: Remove main PC board.
A) Tag / disconnect all wire harnesses from PC board.
B) Cut cable tie, then disconnect pressure sensor tubing.
C) Remove two nuts and PC board / bracket.
Step 2: Install main PC board.
A) Install PC board / bracket, secure with two nuts.
B) Connect pressure sensor tubing, then secure with
high-temperature cable tie.
C) Connect all wire harnesses to PC board.
D) Unplug the power cord.
E) Move switch 2 on the SW1 block to ON.
F) Reconnect power cord.
Step 3: Configure the main PC board.
A) Adjust the PC board configuration by following
the prompts on the display panel.
Use the < + > < - > buttons to adjust settings.
Press the < P > button when finished.
CHAMBER DIAMETER:
M9/M9D ...................... 9 INCH
M11/M11D ..................11 INCH
FEATURE SET:
M9/M11 ....................... FULLY FEATURED
M9D/M11D .................DEFEATURED
Pressure Sensor Tubing
Step 4: Return to normal operating mode.
A) Disconnect power cord.
B) Move switch #2 (SW1 block) to OFF.
C) Reconnect power cord.