Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Error Code:
C633 - continued
C633: HEATUP MODE
PRESSURE LOW
ITEMS NOT STERILE
UNPLUG / REPLUG UNIT
This will occur if chamber pressure
does not reach 1 psi within one
minute after air valve closes; or if
chamber pressure does not reach
4.93 psi within 25 minutes after
starting the HEAT UP mode.
If the unit is building temperature....
Temperature sensor malfunctioning. Check for residue build up on temperature
sensor probe.
Clean sensor probe with Speed-Clean,
distilled water, and abrasive pad.
Check that nothing is touching or crowding
the temperature sensor.
(Example: trays, pouches, cassettes, etc.)
Reposition items so nothing touches or
crowds the temperature sensor.
Perform Temperature Sensor
Resistance Test. Allow sensor to cool to
room temperature before testing.
Acceptable range 1.07-1.1K Ω. Refer to:
Temperature Sensor Resistance Test
Replace temperature sensor.
Perform Temperature Sensor Supply
Voltage Test. Disconnect sensor
harness from J12 on PC board.
On pins 1 & 2 acceptable range is
4.0 - 6.0 VDC. Refer to:
Temperature Sensor Supply Voltage Test
Replace Main PC board.
Pressure Leaks. Check for loose / worn door and dam
gaskets or water leaking around door.
Verify door and dam gaskets are installed
properly. (Replace gasket kit 1x year)
Clean / Replace gaskets.
Replacement fill / vent valve solenoid
requires rectifier harness for DC coil.
Install rectifier harness.
Check for voltage flow through rectifier
harness. Refer to:
Fill / Vent Valve Supply Voltage Test
Replace rectifier harness.
Leaking or defective vent valve.
(Check for water leaking from condensing
coil spout)
Perform: Vent Valve Test
Clean / Replace vent valve.
Leaking or defective fill valve.
Check for water leaking back into reservoir thru
the fill line.
(Look for bubbles coming from bottom of
reservoir) Perform: Fill Valve Test
Clean / Replace fill valve.
Leaking or defective air valve.
Check for excessive steam coming
thru condensing coil spout.
NOTE: During the HEAT & VENT
modes, it is normal for steam to be
exhausted thru the spout.
Clean / Replace air valve.
Leaking or defective Pressure Relief
Valve. Check for water / steam leakage
from beneath the rear of the sterilizer.
Perform: Pressure Relief Valve Test
Replace pressure relief valve.
Pressure transducer tubing leaking.
Check for steam leakage at pressure
transducer tubing connections.
Secure pressure transducer tubing
connections with high temperature cable ties.
Check all plumbing fitting connections for
leakage.
Tighten or replace fittings.