Troubleshooting Chart - continued
Problem Display / Symptom Cause Check Action
Error Code:
C643 - continued
C643: HEATUP MODE
PRESSURE HIGH
ITEMS NOT STERILE
PUSH STOP TO RESTART
PC Board malfunctioning.
Perform: Heating Element Supply
Voltage Test (With harness attached
to leads should have line voltage in
test mode)
Replace Main PC board.
Perform: Main PC Board Relay Test
(Check TP1 & TP2 for an acceptable range:
10 to 14 VDC)
Perform: Main PC Board Pressure
Transducer Voltage Test
(Acceptable range: 4.0 to 6.0 VDC)
Blockage or malfunction in
VistaCool System if used.
Check for debris or blockage in the tubing /
fittings to the VistaCool System.
Clean / Replace as required.
Check operation of In-Iine Thermal Sensor. Replace Thermal Regulator Wax Motor
or Thermal Sensor.
Check for weep hole in the manifold lid and
threaded area on the tank of the VistaCool System.
Drill hole in VistaCool manifold and threaded
area on the tank.
Error Code:
C644
.
C644: STERILIZE MODE
PRESSURE HIGH
ITEMS NOT STERILE
PUSH STOP TO RESTART
Sterilizer Overloaded. Check the load size. (Refer to User Guide) Reduce load size.
Temperature sensor malfunctioning. Check for residue build up on temperature
sensor probe.
Clean sensor probe with Speed-Clean,
distilled water, and abrasive pad.
Check that nothing is touching or crowding
the temperature sensor.
(Example: trays, pouches, cassettes, etc.)
Reposition items so nothing touches or
crowds the temperature sensor.
Perform Temperature Sensor
Resistance Test. Allow sensor to cool to
room temperature before testing.
Acceptable range 1.07-1.1K Ω.
Refer to:
Temperature Sensor Resistance Test
Replace temperature sensor.
Perform Temperature Sensor Supply
Voltage Test. Disconnect sensor
harness from J12 on PC board.
On pins 1 & 2 acceptable range is
4.0 - 6.0 VDC. Refer to:
Temperature Sensor Supply Voltage Test
Replace Main PC board.
Pressure tubing at incorrect angle. Check tubing for restrictions and
ensure tubing is angled down and
away from the PC Board.
Adjust pressure tubing to correct angle.
PC Board malfunctioning.
Perform: Heating Element Supply
Voltage Test (With harness attached
to leads should have line voltage in
test mode)
Replace Main PC board.
Perform: Main PC Board Relay Test
(Check TP1 & TP2 for an acceptable range: 10
to 14 VDC)
Perform: Main PC Board Pressure
Transducer Voltage Test
(Acceptable range: 4.0 to 6.0 VDC)