EasyManua.ls Logo

Milestone DMA-80 evo - 5.15.2.3 Blind values

Milestone DMA-80 evo
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Milestone Srl MA213-003 – DMA-80 EVO – User Manual
103
5.15.2.3 Blind values
Automatic blank value cycles: here it is possible to set the values for the
automatic cleaning of the system after a measurement with a certain ng value of
mercury. (No positions of the plate will be to set since the sample boat holder will
remain into the catalyst.)
In the example: In case of measurement resulting with Hg value corresponding
to 100 ng or more, the blank cycle starts up automatically to get a lower blank <
1 ng for a maximum of 5 times.
You can also set different ng 0-10000, blank cycles 0-100 and at the end of the
cycle it’s possible to select to “Continue” or “abort” the next run.
NOTE: for deep cleaning of boats and instrument refer to procedures described
on chapter 5.11 and 8.3
Blind value method: here are the values of the blind value used for autoblank
cycle.
The blind value can be also used manually by the user for cleaning the machine.
It can be set in the measurement page.

Table of Contents

Other manuals for Milestone DMA-80 evo