Milestone Srl MA213-003 – DMA-80 EVO – User Manual
109
5.15.2.6 Signaling
Signal error: if enabled, a sound is emitted by the controller after an error
message
Signal termination: if enabled, a sound is emitted by the controller after the
analysis is completed
Sound volume and duration: you can choose the volume and time of the sound
5.15.2.7 Maintenance
Exchange of amalgamator: A software message will inform the user that it is
time to replace the amalgamator when the currently cycles have reached the set
limit (5000).
The limit value should be decided on basis of application since the lifetime of
amalgamator strictly depends by it.