Milestone Srl MA213-003 – DMA-80 EVO – User Manual
110
Milestone suggest to replace both amalgamator and catalyst at the same time
when one of the two reached the end of lifetime
Exchange of catalyst: A software message will inform the user that it is time to
replace the catalyst tube when the currently cycles have reached the set limit
(5000).
The limit value should be decided on basis of application since the lifetime of
catalyst tube strictly depends by it
Milestone suggest to replace both amalgamator and catalyst at the same time
when one of the two reached the end of lifetime
Switch off all heating elements: This function is useful during the catalyst
replacement: All heating elements stop to work but the unit keeps ON. In this
way the fans cool more quickly the tube.
Turning back, the heating elements will start to work again.
5.15.2.8 Equipment
Additional sample disk: It is possible to add an addition disc to the instrument
once the first one is completed. If activated, this function allows to prepare a list
of 80 sample (40+40) and add the additional plate when all sample of the first
plate have been measured. The software will inform the user when it is the time
to change the plate.