Milestone Srl MA213-003 – DMA-80 EVO – User Manual
87
Settings/Analysis, the absorbance is calculated like height of the peak of signal
or area.
The value refers to the first signal within the limit of the absorbance of the cell. If
the limit is exceeded, then the next signal in the next cell will be considered.
ng (only available in sub-section “Result”): Absolute quantity of Hg in the sample
(calculated from active calibration).
Hg concentration (only available in sub-section “Result”): Concentration of Hg
in the sample (calculated from Hg quantity and sample weight).
Different unit are available, but the results will be calculated and shown
considering the unit of sample amount.
Cal-Factor (only available in sub-section “Result”) : here is reported the value
calculated by the “theoretical value” divided by “measured value” of reference
sample. See chapter 5.12.4
Dilution (only available in sub-section “Result”): This field is available when you
set a method with “ml” like unit by clicking on cal. factor field.
The final result can be influenced by the dilution factor. The dilution factor can
correct the dilution of an original sample by the digestion.
Example: 1,605 g of a sample has been diluted in 45 ml. Dilution factor: 28,037.
When using the factor, the result is calculated with reference to the original
sample and not to the measured solution.
The input field dilution disappears by clicking another field. The dilution is
displayed under cal factor (here 333,33).