EasyManuals Logo

Mindray A7 Service Manual

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
Preparation - Additional Material Required Installation Guide
2 - 2 046-006272-00 A7â„¢ Service Manual
2.1 Preparation - Additional Material Required
The following additional material are required before installation. The customer is responsible for
supplying this material. Missing items may result in delays, incomplete installations, and/or additional
service visits.
• Compatible emergency O2, N2O, and AIR cylinders
• Agent vaporizers and key fillers (if not purchased with the anesthesia system)
• Liquid agent medication
• CO2 absorbent Pre-Paks or loose fill
• Active O2, N2O, and AIR lines (280 to 600 kPa (40 to 87 psi))
• Dropdown hoses for ceiling-mounted medical gas utilities that are compatible with quick-
disconnect hoses (if not purchased with the anesthesia system)
• Negative pressure source, negative pressure source connection related hoses(if not purchased
with the anesthesia system)

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals