Disassemble the Assemblies Repair and Disassembly
6 - 66 046-006272-00 A7™ Service Manual
6.2.16 Remove Anesthesia Module Inlet Pipeline Assembly
1. Open the service door and refer to“Open the Service Door” on page 6-3.
2. Remove the rear panel and refer to“Remove the Rear Cover Plate Assembly” on page 6-29..
3. Unplug the tubes from the anesthesia module and two-way connector.
FIGURE 6-144
4. Unscrew the two screws to remove the anesthesia module inlet pipeline assembly.
FIGURE 6-145
Cut the cable tie used to fix the filter and disconnect the tube from the filter to remove the filter.