Functional Tests Installation Guide
2 - 30 046-006272-00 A7â„¢ Service Manual
2.3.3 Display Setup Check
Touch Continue to advance to the Standby screen display. For the A7, check that the screen displays
the main screen, check that all ventilation modes are there, as shown in the figure below.
FIGURE 2-43 Display Setup Check, A7
2.3.4 Calibrate the AG Module
Prepare the following before doing the external and Internal AG module calibration:
• Gas cylinder, with a certain standard gas or mixture gas. Gas concentration should meet the
following requirements: AA≥1.5%, CO2≥1.5%, N2O≥40%, O2≥40%, of which AA represents
an anesthetic agent. a/c≤0.01 (a is the gas absolute concentration accuracy; c is the gas
concentration).
• T-shape connector
• Tubing
Follow this procedure to perform a calibration:
The BFCS fails to retract
during automatic backup
flow control test.
The stepper motor of BFCS is faulty.
1. Re-start the machine.
2. Replace the parts of backup
flow control system.
The three-way valve is
faulty.
When the three-way valve is switched to
BFCS limb during BFCS automatic test, the
measured flow of total flow meter is less
than 0.2L/min.
1. Check if the gas supply
connection and supply gas
pressure are normal.
2. Replace the three-way valve.
Failure description Possible cause Recommended action

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals