A7™ Service Manual 046-006272-00 2 - 31
Installation Guide Functional Tests
1. Connect the test system as follows.
FIGURE 2-44
NOTE: When calibrating the internal AG module, open the back cover of the
machine, disconnect the tubing as shown below and connect the
calibration gas to the gas inlet.
FIGURE 2-45
Relief valve
Gas cylinder
Tubing
AG module
Open to the air
Disconnect the tubing here Gas inlet