A7â„¢ Service Manual 046-006272-00 5 - 81
Repair and Troubleshooting Software Update and Software Configuration Activation
5.6 Software Update and Software Configuration
Activation
NOTE: Software upgrade may by required when replacing the CPU Board (P/N:
801-0631-00026-00), the Ventilator Control Board (P/N: 801-0631-
00027-00), the Power Board (P/N: 115-018145-00) or the EFCS Board (P/
N: 115-018150-00).
1. Connect the Ethernet port of the PC to the Ethernet port of the A7 using the Ethernet Crossover
Cable.
2. Before running Mindray Patient Monitor System Update Tool, verify that the IP address of the PC
is set to 192.168.23.1., and the Subnet mask is set to 255.255.255.0. To check and set the IP
address on the PC follow these instructions.
a. On the PC click Start, Settings and then Network Connections.
b. Right click Local Area Connection and then left click Properties.
c. Scroll down to Internet Protocol (TCP/IP), click on it and then click Properties.
d. Click the radio button for "Use the following IP address:" Set the IP address and Subnet mask
and then click OK.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals