A7â„¢ Service Manual 046-006272-00 2 - 19
Installation Guide Assembly
NOTE: The barometric pressure may differ from the calibration pressure of the
anesthetic vaporizer. This may cause an inaccurate output of the
anesthetic agent. The operator should continuously monitor the
concentration of anesthetic agent during system use.
2.2.3.1 Mount the Vaporizer(s)
1. Mount the vaporizer onto the manifold.
2. Push and turn the locking lever clockwise to lock the vaporizer in position.
3. Ensure that the top of the vaporizer is horizontal. If not, remove the vaporizer and reinstall it.
4. When reinstalling the vaporizer, lift each vaporizer straight up off the manifold rather than
pulling forward. Do not rotate the vaporizer on the manifold.
5. If a vaporizer unintentionally lifts off the manifold, install it again and complete steps 1 through
3. If the vaporizer lifts off a second time, do not use the system.
NOTE: A Desflurane vaporizer may be mounted similarly as other vaporizers,
but may require a power cord. For more detailed instructions on
installation and proper use, refer to the specific manufacturer's
Instructions for Use of the Desflurane vaporizer.
2.2.3.2 Fill the Vaporizer
NOTE: The A7 should use vaporizers with Selectatec mounting system that are
compliant to ISO 8835-4. Refer to the vaporizer manufacturer's
Instructions For Use for filling or draining the vaporizer and other
information.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded label. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
2.2.3.3 Drain the Vaporizer
WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous
chemical and follow local regulations for proper disposal.
NOTE: The A7 should use vaporizers with Selectatec mounting system that are
compliant to ISO 8835-4. Refer to the vaporizer manufacturer's
Instructions For Use for filling or draining the vaporizer.
2.2.4 Monitoring Products Mounting and Electrical Connection
(if available)
1. Mount the monitor (if available) according to the manufacturer's monitor assembly instructions.
NOTE: Use of other monitors and mounting hardware is the responsibility of
the installer.
2. After mounting a monitor to the A7, connect it to one of the AC outlets located on the rear of
the A7.
3. Turn on each monitor one at a time. Verify that the circuit breaker holds without tripping.
4. Dress each line cord neatly along the side of the anesthesia machine or tucked inside the
monitor arm. Ethernet and power cables can be routed through the clips on the rear door of
the A7.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals