EasyManuals Logo

Mindray A7 Service Manual

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #211 background imageLoading...
Page #211 background image
A7â„¢ Service Manual 046-006272-00 3 - 51
Periodic Maintenance Electrical Tests
3.17.3 Electrical Safety Inspection Form
For periodically performance, all the test items included in the ELECTRICAL SAFETY INSPECTION
FORM shall be performed. The following table specifies test items to be performed after the
equipment is repaired with main unit disassembled.
Location: Technician:
Equipment: Control Number:
Manufacturer: Model: SN:
Measurement equipment /SN: Date of Calibration:
INSPECTION AND TESTING Pass/Fail Limit
1 Auxiliary mains socket outlets
2 Protective Earth Resistance Ω Max 0.1 Ω
3 Earth Leakage
Normal
condition(NC)
____μA
Max:
NC: 300μA
SFC: 1000μA
Single Fault
condition(SFC)
____μA
TABLE 3-1
When neither power supply PCBA, transformer nor
patient electrically-connected PCBA is repaired or
replaced
Test items: 1, 2
When power supply PCBA or transformer is
repaired or replaced
Test items: 1, 2, 3
TABLE 3-2

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals