Disassemble the Breathing System Repair and Disassembly
6 - 76 046-006272-00 A7â„¢ Service Manual
6.3.5 Remove the CO2 Bypass Assembly
1. Remove the absorbent canister as per section 6.3.5.
2. Press inward the fasteners on both sides and the CO2 bypass assembly will drop down for
removal.
FIGURE 6-162
If it is necessary to replace or re-install the bypass trigger plate sealing cushion (049-000142-00), pay
attention to the matching between the interior convex rib of the bypass trigger plate sealing cushion
and the groove of the bypass trigger plate when assembling the bypass trigger plate sealing cushion
onto the bypass trigger plate, as shown below.
FIGURE 6-163
Make sure that the sealing cushion is correctly installed. The comparison between correct installation
and incorrect installation is shown below.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals