EasyManuals Logo

Mindray A7 Service Manual

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
A7â„¢ Service Manual 046-006272-00 2 - 49
Installation Guide Alarms and Fail safe Functions
2.7.4 Peak Pressure Alarms Test
1. Set the PEAK low alarm limit to the lowest setting.
2. Set the PEAK high alarm limit set point about 5 to 8 digits below the Peak Pressure displayed on
the screen.
3. Verify the following (high) peak pressure alarms activate:
• Paw Too High message appears on the screen.
• An alarm tone sounds.
• Inspiration ends and expiration begins as the pressure meets the high alarm limit.
4. Set the PEAK high alarm limit to the highest setting.
5. Verify the Paw Too High message disappears.
6. Set the PEAK low alarm limit to 70 (cmH2O).
7. Verify the following (low) peak pressure alarms activate:
• Paw Too Low message appears on the screen.
• An alarm tone sounds.
8. Set the PEAK low alarm limit to the lowest setting.
9. Verify the Paw Too Low message disappears.
2.7.5 Minute Volume Alarm Test
1. Set the MV High alarm limit to the highest setting and then set the MV Low alarm limit to the
highest setting.
2. Verify the following alarms activate:
• MV Too Low message appears on the screen.
• An alarm tone sounds.
3. Set the MV Low alarm limit to the lowest setting.
4. Verify the MV Too Low message disappears.
5. Set the MV High alarm limit to the lowest setting.
6. Verify the following alarms activate:
• MV Too High message appears on the screen.
• An alarm tone sounds.
7. Set the MV High alarm limit to the highest setting.
8. Verify that the MV Too High message disappears.
9. Set the mechanical Auto/Manual switch to MANUAL.
10. Set all fresh gas flows to 0.
11. Press End Case button to enter the standby mode.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals