Performance Verification Installation Guide
2 - 44 046-006272-00 A7â„¢ Service Manual
2.6.4 VCV Adult Ventilation Mode Test
1. Set the Fresh Gas Control to Direct Flow Mode.
2. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
3. Set the mechanical Auto/Manual switch to AUTO.
4. Set the ventilator controls to:
5. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp. rate and minute
volume values appear on the screen.
6. Verify the Tidal Volume display on the Vent Tester is within 7% (±35 mL) of the set value within
approximately 1 minute from the start of ventilation.
7. Verify the Tidal Volume display is within 9% (±45 mL) of the set value within approximately 1
minute from the start of ventilation.
8. Verify the PEEP on the display and on that the Vent Tester is between 0 and 4 cmH2O.
9. Verify the measured O2 concentration is at least 97% after 5 minutes.
10. Set the AIR flow to 3 L/min and set the N2O and O2 flow rates to minimum flow.
11. Verify the measured O2 concentration is 21% ±3% vol. % after 5 minutes.
2.6.5 VCV Adult Ventilation Mode Test 2
1. Set the O2 flow to 10 L/min and set the N2O and Air flow rates to 0 flow.
2. Set the mechanical Auto/Manual switch to AUTO.
3. Set the ventilator controls to:
Ventilator Controls Ventilator Settings
Ventilation Mode VCV
Vt 500
Rate 4
I:E 1:3
Tpause 30
PEEP Off
Plimit 50