Functional Tests Periodic Maintenance
3 - 26 046-006272-00 A7â„¢ Service Manual
FIGURE 3-28 Calibration Front Control Panel - Flow Setting
c. Gas Settings: Press the Setup button, select Setting->ENTER->Gas Settings->MODIFY->Gas
Type->O2.
d. Correction mode: Press the Setup button, select Setting->ENTER->Correction Mode-
>MODIFY->Correction Mode->ATP..
FIGURE 3-29 Calibration Front Control Panel - Gas Settings
e. Select BACK->BACK-> BACK.