System Calibration Calibration
4 - 46 046-006272-00 A7â„¢ Service Manual
4.3.4.1 Zeroing Procedures
Follow these steps to zero pressure and flow sensors.
1. Select Setup-> Service-> Calibration-> Zero Sensors to access the screen shown below.
FIGURE 4-59
2. Select Begin to access the zeroing screen as show below. During the zeroing, you can select
Cancel to cancel the zeroing.