System Calibration Calibration
4 - 50 046-006272-00 A7â„¢ Service Manual
4.3.5.1 Zeroing Procedures
Perform the following steps to zero the EFCS.
1. Select Setup > General> Zero Flow Meters to enter the screen shown below. Select Begin to
start zeroing.
FIGURE 4-64
2. The zeroing screen shown below is displayed when Begin is selected. During the zeroing, you
can select Cancel to cancel the zeroing.