EasyManua.ls Logo

Mindray A7

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Calibration Calibration
4 - 54 046-006272-00 A7™ Service Manual
FIGURE 4-70
NOTE: Realtime fresh gas flow is displayed on the [Calibrate Total Flow Sensor]
screen. Normally, only N2O and total flow values are greater than 0, and
N2O and total flow values ascend with the increase of calibration
progress. Air and O2 values should be 0 or less than 0 all the time.
4. The following screens are displayed after calibration is completed.
The screen shown below is displayed if total flow sensor calibration fails. Select [Try Again] to
calibrate again or [Done] to exit calibration.
FIGURE 4-71
5. The screen shown below is displayed after successful total flow sensor calibration. Select [Done]
to exit calibration.

Table of Contents

Other manuals for Mindray A7

Related product manuals