A7™ Service Manual 046-006272-00 4 - 59
Calibration System Calibration
FIGURE 4-77
4. Start the system, turn on the cylinder gas supply (turn off the pipeline gas supply), and adjust
the fresh gas flow to 1 L/min.
5. Remove the self-locking acorn nut at the head of the regulator. Install the cylinder.
FIGURE 4-78
Self-locking nut