EasyManua.ls Logo

Mindray A7

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A7™ Service Manual 046-006272-00 4 - 61
Calibration System Calibration
NOTE: Different pressure meters may require different adapters.
FIGURE 4-81
FIGURE 4-82
2. Disconnect the pipeline on the back end of the back pressure valve.
3. Connect to O2 source, and adjust O2 needle valve, while observing the flowmeter on the
system interface of A7, to set the O2 flow rate to 5±0.1L/min.
4. Loose the nut of the back pressure valve.
5. Adjust the bolt of the back pressure valve, while observing the displaying pressure value of
Anesthesia machine calibration device, to set the pressure value to 237 ± 10 cmH2O.
6. Lock the Nut, as shown in the following figure.
Pressure monitor
port
3140-08-00 Y piece
Connect to the
pressure monitor
port of Anesthesia
machine calibration
3106-10-00 adapter
connector

Table of Contents

Other manuals for Mindray A7

Related product manuals