Technical Alarms Check Repair and Troubleshooting
5 - 18 046-006272-00 A7â„¢ Service Manual
Auto Ventilation is
Non-Functional
High
Manual Only. But the
Auto/Manual switch
is in Auto position.
1. Place the Auto/Manual switch in Manual position.
ACGO Failure Med
The status of the
ACGO switch is
incorrect.
1. Check whether the drive gas pressure is within the normal range. If
not, ensure that the drive gas pressure is within the normal range.
Then turn on or off the ACGO switch.
2. If the problem persists, replace the ACGO switch.
3. If the problem persists, replace the CPU board.
4. If the problem persists, replace the VCM
Electronic ACGO
Undetected
Low
No electronic ACGO
is detected.
1. Check whether an electronic ACGO module is configured for the
machine.
2. If so, check the connections between the ACGO module and the
CPU board, and between the ACGO module and the VCM.
3. Replace the electronic ACGO module when necessary.
4. Replace the CPU board or VCM when necessary.
Message Priority Cause Solution

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals