Reviewing Sample Results
7-33
is less than 0.2 or non-numeric (***), the RBC histogram is not adjustable.
The first two discriminators of the PLT histogram are adjustable. Note that if the PLT result is
less than 10 or non-numeric (***), the PLT histogram is not adjustable.
For example, to move the third discriminator of the following WBC histogram, follow the
procedure below to do so.
1. Press [ENTER] and the discriminator will become adjustable. See Figure 7-50;
Figure 7-50 WBC histogram with adjustable discriminators
2. Press [] or [] as needed to select the WBC histogram, as Figure 7-51 shows;
Figure 7-51 Adjusting discriminator (1)
3. Press [F3] to select the third discriminator, as Figure 7-52 shows;
Figure 7-52 Adjusting discriminator (2)